Advanced Photon Source at Argonne National Laboratory   APS USAXS instrument
USAXS instrument at the Advanced Photon Source,
X-ray science division, beamline 9ID-C

Advanced Photon Source

A U.S. Department of Energy, Office of Science,
Office of Basic Energy Sciences national synchrotron x-ray research facility

 
 
Argonne Home > Advanced Photon Source > USAXS >

PDF version

Schedule
Staff web pages:
      Jan Ilavsky, inst. scientist
Documentation

      Glassy Carbon int. stnd.
      Example USAXS data
      Getting beam time
      Instrument Geometries
      USAXS imaging
      Overview
      Posters
      Sample Holders
      Select publications
      User publications
      Shipping Instructions
Live USAXS data
Software by Jan Ilavsky:
      Indra (reduction)
      Irena (analysis)
      Nika (2D data reduction)
      Cromer-Liberman AtFF
      Clementine (kinetic rate analysis)
      other macros
Useful WWW links


APS 9ID USAXS/SAXS/WAXS
Ultra-Small-Angle X-ray Scattering Facility

e-mail: usaxs@aps.anl.gov, instrument scientist: Jan Ilavsky, 630-252-0866, ilavsky@aps.anl.gov and Ivan Kuzmenko, 630-252-0327, kuzmenko@aps.anl.gov

USAXS Absolute Intensity

Pete R. Jemian, jemian@anl.gov

September 2005

from the 1990 Northwestern University PhD thesis:
http://www.jemian.org/pjthesis.pdf

Conversion of USAXS scattering data into differential cross-section in units of cm$^{-1}$ involves measuring the ratio of the number of photons scattered per second into unit solid angle to the number of photons in the incident beam.

The as-measured slit-smeared intensity profile (in some arbitrary units),

\begin{displaymath}
\tilde I(Q) = I_0 \ \Omega \ t \ T_s \ {{d\tilde\Sigma} \over {d\Omega}}(Q)
\end{displaymath} (1)

is measured by a detector subtending a solid angle, $\Omega$, with the sample. $I_0$ is the apparent source intensity incident on the sample, with identical units as $I(Q)$ and identical efficiency of detection. $d\tilde\Sigma(Q) / d\Omega$ is the slit-smeared differential scattering cross-section per unit volume per unit solid angle. The measured transmission is given by $T_s = e^{-\mu t}$ where $\mu$ is the linear absorption coefficient of the sample and t is the sample thickness.

With the USAXS instrument, the same detector that measures $I(Q)$ is used to measure $I_0$ with the sample removed from the instrument, the detector efficiency cancels and primary conversion of the measured intensity into units of cm2cm-3sr-1 involves only the measurement of t, T, $\Omega$, and $I_0$. That is, the slit-smeared SAXS in absolute units measured by the USAXS instrument is

\begin{displaymath}
{{d\tilde\Sigma} \over {d\Omega}}(Q) =
{
{R_s(Q)/T_s - R_b(Q)}
\over
{R_b(0) \ \Delta\theta_w \ \Delta\theta_l \ t}
}
\end{displaymath} (2)

where $ \Delta\theta_w$ and $ \Delta\theta_l$, which define the detector solid angle $\Omega$, are the angular width and length of the beam intercepted by the detector.

The ratio, $R(Q)$, of detector counts, $D$, in time, $t$, to monitor counts, $M$, corrected for detector range dark current, $d_r$, and adjusted for detector range amplification, $A_r$, measured at each different Q, is

\begin{displaymath}
R(Q) =
{
A_r \left[D(Q) - \tau d_r\right]
\over
M(Q)
}.
\end{displaymath} (3)

The intensity ratio, $R(Q)$, is a superposition of the small-angle scattering from the sample, small-angle scattering from any windows between the collimating and analyzing crystals, convolution of the X-ray source profile with the rocking curves of the monochromator and USAXS optics, and parasitic background scattering of the X-ray beam in the air path between the sample position and the detector. The source convolution will be shown later to be negligible. At $Q=0$, the USAXS analyzer crystal conducts the full intensity of the unscattered beam ($I_0$, the intensity apparent at the detector) so the sample transmission,
\begin{displaymath}
T_s = {R_s(0) \over R_b(0)},
\end{displaymath} (4)

where $R_s(Q)$ and $R_b(Q)$ are the curves measured for the sample and blank beam, respectively, normalized to a constant monitor count rate. The blank beam is a scan which is the same as a SAXS scan in every regard except that the sample is removed.

It may be assumed that all effects except the scattering from the sample and a small additional background will be found in both $R_s$ and $R_b$. The instrumentally-smeared small-angle scattering plus small additional background in arbitrary units is then separated by scaling the sample data by its transmission factor and subtracting the empty beam, as

\begin{displaymath}
\tilde I(Q) = R_s(Q)/T_s - R_b(Q).
\end{displaymath} (5)

In a SAXS camera that utilizes geometric collimation with a concomitantly small illuminated area, e.g. (Wignall & Bates, 1987), the solid angle is defined by the area of the beam on the detector element, $\Delta a$, and the distance between the sample and detector, $SDD$. However, in the USAXS instrument, the angle in the scanning direction is more highly-collimated by the rocking curve of the crystal optics. Therefore, $ \Delta\theta_w$ is measured from the experiment as the full width at half maximum of the empty beam, $R_b(Q)$. With no crystal optics in the slit-length direction, $ \Delta\theta_l$ is defined in the geometric sense. For a large illuminated sample area defined by the width and length of the source slits, where $A = W_S \times L_S$, and a detector diameter of $L_D$, where $L_S < L_D$,

\begin{displaymath}
\Delta\theta_l =
{L_D + L_S \over 2SDD} + {L_D - L_S \over 2SDD} = {L_D \over SDD}.
\end{displaymath} (6)

The detector diameter (diameter of the active area of the unbiased PIN photodiode) used in the UNICAT USAXS instrument is 5.5 mm and the sample-to-detector distance is typically 300 mm. Since the sample-to-detector distance need be known to only a precision of 5 mm or so, the additional X-ray path distance due to the mutiple reflections in the analyzer is negligible with respect to this precision. Thus it is evident that although primary intensity calibration usually requires special equipment and a standard sample, and can be difficult and time-consuming to perform, the USAXS instrument requires no additional beam time for calibration since all of the parameters are available in the normal course of performing scattering measurements.



Pete R. Jemian, jemian@anl.gov, 2005-09-09

        